Salera preselect hip joint
Delivery time approx. 5 working days
Product information "Salera preselect hip joint"
The Salera preselect is a 3-axis hip joint that actively allows limited circular movement. Thanks tot he sophisticated design, the lower part oft he hip joint allows a defined rotational movement (12.5° internal or external rotation) despite complete blocking of adduction. The extension and flexion are limited by the existing stop surfaces.
The step length can be individually adjusted and released by using interchangeable wedges.Due to the design change of the Salera preselect, the sliding surface has been enlarged.This leads to a reduction in the coefficient of friction and thereby reduces wear. By releasing the lock, it is possible to sit down. More information on step processing with the SALERA preselect hip joint and the preselect function of the lock can be found here.
Advantages:
- Physiological sitting is possible
- Allows rotational movement
- Interchangeable wedges allow individual adjustment of the step length
- Greater safety thanks to the preselect function of the lock
Use:
- For HKAFO
- lesions between L5 and L3 (the patient can still actively lift the pelvis)
Note:
The quadratus lumborum muscle must retain its function. This is because it triggers the pelvic elevation and thus represents the last option to support the SALERA preselect hip joint through the iliac crest profiling when lifting the free leg.
Size chart:
- SALERA preselect small: up to a body weight of 25 kg
- SALERA preselect medium: up to a body weight of 40 kg
- SALERA preselect large: up to a body weight of 70 kg
.
Scope of delivery:
- including pelvic fork: small: 6 x 50 x 400 mm / medium: 8 x 50 x 500 mm / large: 10 x 50 x 600 mm
- including extension set: small: 6 x 17 x 270* mm / medium: 6 x 17 x 270* mm / large: 8 x 20 x 325* mm (*Length measured from the middle of the joint screw (joint center) to the end.)
- including interchangeable wedge set
not bent, not welded
Further information on the dimensions of the Salera preselect hip joint can be found here.
Note:
Please note that the following special tools are required to install the Salera preselect hip joints. The tool can be used multiple times.
Locking mechanism assembly-Set for disassembling/assembling the unlocking and locking unit
Additional note:
The hip joints and the pelvic rocker are delivered unprocessed and not welded. As a service manufacturing service, we are happy to assemble and weld the pelvic rocker (Art.-no.: 80.51.050.001). This is done according to measurements or an outline drawing. You can find more information about this here.
We are happy to manufacture a pelvic-wide walking orthosis with Salera preselect hip joints for you in service manufacturing. Please send us the completed " Inquiry service manufacturing orthotic" form. You will then receive a corresponding cost estimate from us. You can find more information about our service manufacturing here.